This instrument is mainly used to measure the oil content in surface water, groundwater, and industrial wastewater. It can also be used to monitor oil in flue gas and solids in the catering industry. This instrument is a professional instrument for measuring the oil content in water, gas, and solids.
Infrared Oil Meter
This instrument is mainly used to measure the oil content in surface water, groundwater, and industrial wastewater. It can also be used to monitor oil in flue gas and solids in the catering industry. This instrument is a professional instrument for measuring the oil content in water, gas, and solids.
Instrument Features:
◆The instrument comes with an embedded 10.8-inch win10 system tablet computer with full HD display;
◇Built-in professional oil measurement software, the software spectrum is clear and can distinguish various interference phenomena;
◆The software has functions such as infrared spectrum analysis, spectrum storage, data query, statistical calculation, and printing;
◇Opto-mechanical and electrical integration design and electrically modulated light source improve instrument reliability;
◆High accuracy: the error of each measurement does not exceed ±1.5%;
◇Baseline drift is less than 0.005AU/60min;
◆Has the function of direct reading of non-dispersion measurement results without conversion;
Fundamental:
According to the national standards GB16488-1996 and HJ637-2018, use tetrachlorethylene to extract oil substances in water, then use anhydrous sodium sulfate to isolate impurities and moisture, separate the solution of oil and tetrachlorethylene, and measure the total oil content of the extract.
| Performance Parameters | |
| Basic Measuring Range | 0~120mg/L |
| Lowest Detection Limit | 0.001mg/L (Water Sample Concentration) |
| Linear Correlation Coefficient | r﹥0.999 |
| Maximum Measured Concentration | 100% Oil |
| Wavenumber Range | 3400cm-1~2400cm-1 (i.e. 2941nm~4167nm) |
| Repeatability | The 30-40 Mg/L Oil Standard Sample Was Measured 11 Times With Rsd ≤ 0.6% Or The 0.5 Mg/L Water Sample With Rsd ≤ 2%. The Optical System And Electrical System Of The Instrument Are Self-Contained And Highly Integrated, Thereby Improving The Reliability And Reliability Of The Instrument. Repeatability. |
| Wavenumber Accuracy And Repeatability | ±0.5cm-1. Using Positioning Calibration, Oil Samples Above 30mg/L Are Automatically Calibrated At 2930cm-1, 2960cm-1, And 3030cm-1, Eliminating The Influence Of Artificial Positioning. |
| Correction Coefficient Accuracy | Use Tetrachlorethylene As The Reference Solution And Use a 4cm Cuvette To Measure 0mg/L, 5mg/L, 20mg/L, 50mg/L And 100mg/L Petroleum Standard Solutions Respectively, With An Error Of ±5%. |
| Absorbance Range | 0.0000~3.0000Au (i.e. Transmittance 100%~1%T) |
| Method Detection Limit | When The Sample Volume Is 1000ml And The Extraction Liquid Volume Is 25ml, The Detection Limit Is 0.01mg/L; |
| Instrument Detection Limit | Dl≤0.04mg/L (3 Times Sd Of 11 Times Of Tetrachlorethylene Blank Solution Measurement); |
| Scan Speed | Full Spectrum Scan, 30 Seconds/Time; Non-Dispersive Infrared Method 2 Seconds |
| Baseline Stability | Mechanical Light-Cutting Technology Is Used To Automatically Adjust The Zero Point (Eliminating The Influence Of Baseline Drift) To Improve The Long-Term Stability Of The Signal; |
| Solvent Selection | Tetrachlorethylene, S-316, Trichlorotrifluoroethane And Other Non-Carbon Chlorine Organic Solvents Can Be Used As Extraction Agents |
| Physical parameters | |
| Host weight | 12kg |
| temperature humidity | 15-30℃ 20%-80% (no condensation) |
| Power supply | (220±22)V、(50±1)Hz、50W |
| Dimensions | 480*340*160mm |
| Light source system | The light source has a service life of more than 50,000 hours and uses infrared tungsten halogen lamps imported from Germany; |
| Software system | Dual system analysis is adopted to meet the use of tetrachlorethylene as an extraction agent; |